Hi, If you do an ASCII export from the browser (select all the curves you like in the browser and right click-- you should see an export option) in Nanoscope Analysis, there is a "Ramp" check box that allows you to export the Z values. Please note that these Z values are based on the drive voltage since the Multimode does not have a Z sensor
-
Hi Chiara, It looks as if the export is putting out the correct numbers (at least for the file you sent). As you noted, the export is using exponential notation, but if you pay attention to the units in the header line it appears to be correct (i.e. 4e+003 pN = 4 nN): Input1_V_Ex Input1_V_Rt Defl_nm_Ex Defl_nm_Rt Defl_pN_Ex Defl_pN_Rt 3.030396E+00
-
Hi Jörn-Oliver, You can do batch exporting of ramp files to ASCII in either Nanoscope or Nanoscope Analysis by selecting all of the files you want to export (with shift-click or ctrl-click) in the browse view right click to open the contex menu choose Export>ASCII... This will generate a bunch of text files that can be read by Excel or whatever