Hi Sergio, Yes, you can measure the cantilever amplitude in nanometers on the Dimension Nanoman system (and other systems as well). The most common way to do this is to take a force curve while in TappingMode which produces an amplitude vs Z curve. This will show the amplitude change from a constant amplitude to a relatively linear decreases once the
-
Hi Dalia, You are correct that the TMFbIgain and TMFbPgain are the intergral and proportional gains, respectively. Also, the amplitude setpoint is labled TMSetAmplitude. Another way to search in version 6.13 for these items is to right click when you are in properties menu and choose "copy text", and then paste this into an editor like Notepad
-
Hi Nishith, You are correct that the surface delay does not record during the delay, but there are ways to do this using the scripting interface and the strip chart. I would like to see the data that you have so far and get a better understanding of what you would like to do, so I will contact you and we can proceed from there. Thanks, John
-
Hi Nishith, PeakForce QNM on the ICON uses PeakForce Tapping to oscillate the probe at a constant rate while generating force curves, and there is not the option to setup a ramp delay to hold the tip on the sample at the end of the extention for a specified period of time. If this is what you want to do, the standard force curve menu and the "nanoindentation"
-
Hi Catalina, Yes - If you set the Ramp Delay in the Advanced force curve menu, it will still be active in Force Volume mode although that parameter is not shown in that menu. You can verify this by running Force Volume with and without a Ramp Delay, and you will see the time difference. I hope this helps. Please let me know if you have further questions
-
Hi Catalina, The parameter that you are looking for is called "Ramp Delay" in version 5.30. You do not need to upgrade software to access this parameter. When you are in the force menu, it is in the "Scan Mode" panel. If you do not see this panel, go to the Panels pull down menu and choose "Mode". It should be near the
-
I have run MFM in fluid in the past, and it seemed to work. I used the standard probe for MFM (Model MESP) which worked in Tapping in Fluid and used Lift Mode for the MFM signal. You will likely need to adjust the lift height, lift start height, and probably scan slower than normal. I also recall the lower sensitivity that Steve mentioned as well. It
-
Hi Xiaoqian, Following up on Rob's answer, the Cantilever Tune will sweep through a range of frequencies to find the cantilevers resonance frequency. If using AutoTune, it will then setup the oscilation amplitude at the resonance frequency, which will produce the resulting RMS amplitude read by the detector. So, Cantilever Tune will set both the
-
Hi Linda, For your application and in general, it is important to get a good understanding of what the Flatten and Planefilt routines do to the data and how this will effect your results. They are different routines, but as you have experienced there is some similarity between the effect on the image. Flatten modifies the image by adjusting each line