Man, your data looks terrible! This extreme noise is not normal. Also the fact that the Force increases at decreasing Z Sensor, so at increasing Separation, means you have got some serious problems. I have no experience with a D3100 system. To me (as a non-specialist) it seems as if there is some problem with your laser point positioning. Could that
-
Please let me know if you managed to solve your problem.
-
Dear community, I struggle trying to make the Z axis equally sized to the other two axes in the 3D image in Nanoscope Analysis. Depending on the absolute Height differences in the image I need to set "Zaxis Aspect Ratio" to about 0.03 or to 0.14 in order to obtain approximately equally sized axis. According to the NanoScope Analysis help you
-
Hello, tibs, Could you please add your data files so that we can reproduce your problem? Go to your profile at the top right corner of the window and go to "my files" then. Upload your files in form of ZIP-files and add them to your posting via "Insert Media". Best regards, Dietmar
-
What's the problem? Is the live image black or not visible at all? Could it be that you didn't activate "Live image" in the menu?
-
3) Unfortunately, I cannot help you with your last issue; I have to face the same problem: My QNM images often do not agree with the values calculated offline, especially when I use the "Point-&-Shoot" method. This might have to do with the different ways the data is acquired: In Point-&-Shoot mode the tip indentates the sample always
-
Dear Mariam, 1) This was explained here: http://nanoscaleworld.bruker-axs.com/nanoscaleworld/forums/t/1306.aspx . Basically, the answer is that you need to capture an image first (from the Point-&-Shoot menu). Once the image is captured, "Ramp and Capture" becomes enabled. According to http://nanoscaleworld.bruker-axs.com/nanoscaleworld
-
Don't you think that is a bit too little information?
-
I cannot view your images. Did you try to upload them using "My Files" (bottom right of your profile homepage) and to include it using "Insert Media"? Dietmar
-
Dear community, I use MultiMode 8 AFM in ScanAsyst mode using a ScanAsyst Air tip. Recently, I had to realize that the resolution drops very quickly, probably due to tip wear. I used a fresh tip, scanned a few samples and then scanned my first sample again. It looked very different then. In order to check if the reason was tip wear I imaged anotherone