  • Re: Peak force quantitative nanomechanical mapping

    That is generally due to poor deflection sensitivity calibration. If the deflection sensitivity is too large, the change in cantilever bending can be erroneously calculated to be larger than the Z piezo motion. This causes the deformation calculation to return negative numbers and will cause problems...
    Posted to SPM Digest by Bede Pittenger on Thu, May 31 2012