Showing related tags and posts across the entire site.
-
My understanding is the follow ing: Tip fly limit controls when the DSP takes a data point on flat and is directly hooked to micro step size, (e.g. µm step size 2, tip fly limit 2 = data point taken every 4microns on flat. Is this correct? What happens on sidewalls?
-
In addition I find that it's sometimes best to have Tip Bottom Cutoff set to the same height as the tallest feature of interest. While this might not eliminate all artifacts in the post extraction image it should mean that the features you care about are properly resolved. Jason.
-
What is the difference between the "Tip Servo Distance Limit" and the "Lateral Servo Limit"?
-
Is is better to use a larger microstep size and a small tip fly limit or better to use a small microstep size and a large tip fly limit?