Hi Lujie, E-scanner and a Multimode are a good choice for atomic resolution trials. However, there can be many reasons, why you don't get it - starting from badly peeled or not very well fixed sample (particularly with mica, which is a layered material, where layers easily can slip against each other), insufficient vibrational and acoustic isolation
-
Hi Luci, for me this sounds, as you would not need to use a breakout box for that. You should be able to do that nicely in the standard force spectroscopy mode (putting a ramping voltage to the z-piezo is nothing else than performing a z-ramp). If you definitively want to use volts instead of nm, you can choose units "volts" instead of "metric"
-
Hi Alex, interesting question and remarks... Besides the hardware differences (choice of probe, operational mode), the two methods have as well very different probing frequencies. At 1-2kHz probing frequency, a material might have different elastic properties than at 60kHz (viscoelastic materials are typically stiffer at higher frequencies). As we are