Fenny, you might try very lightly wetting a cleaning sample surface with H2O2 hydrogen peroxide, varying the amounts to be able to immerse the tip entirely in the liquid, just the tip, NOT the cantilever, and hold it contacting the peroxide as long as possible. If you can instead, remove the tip and immerse in H202 just to cover the whole lever and
-
A typical error in refurbishing / reinstalling ISA DSP / Interface cards, is the default setting in BIOS for PNP ISA cards setting is incorrect for operating the ISA cards under WIN NT ( poss later Operating systems too ). If I remember correctly, you need to Disable PNP in the BIOS for ISA bus slots, ( PNP is Plug N Play ). Aside from setting up the
-
Hi Nic, Thanks for the kind suggestions. I am doing some involved? KFM ( kelvin voltage imaging, without an EFM tipholder ! that is a trick ) this week and will test these settings out early next week. Mark
-
I tried someone's suggestion to set in the OTHER Controls Panel - to set Engage Setpoint to 1.00 ... Turns out this was set as such all along ( ie not the fix ).
-
Hi, I am resurrecting an ancient NS3 Phase Extended Multimode. ( a wonderful classic instrument ) I loaded up an FIB tip and began normal engage with default Tapping Engage settings in this software, and saw evidence the $200 FIB2-110s tip (spike section) did not survive the engage :-(. For the next month or so I will have to continue with this software
-
Stephan, Thanks for the clarifications. This is quite helpful. Do you have any info as to how to properly wire the analog PFM from witec to implement it with SBOB ( if needed ) and Z board modifications for the NS3 (non A ) ? ( and diagrams from your old work ). Witec was initially ( ?wrongly ) indicating the NS3(non-A) would have difficulty accepting
-
I am curious if anyone has used a Witec 1st generation analog Pulsed Force Microscopy box with a Nanoscope 3 or 3a, and what comments and observations you had with this? I am presently trying to set up an Analog Witec PFM with a Nanoscope 3 system ( non A ) and despite the std Witec (37pin cabled ) DI interface not being (formally) supported by Witec