  • Re: Nanoscope IIIa problem

    Hi Stefan, The Bruker of Brazil stayed to send me a proposal to upgrade the system, but not have sent until now . I'm still waiting . And I had requested a file for Pat to try to recover my old computer, but he did not respond to my request. I need the file " ! CPQ3011.CFG "
    Posted to SPM Digest (Forum) by Matheus on Mon, Dec 17 2012
  • Re: Nanoscope IIIa problem

    Hi , How can I do to run the ECU ? I put those files on a floppy disk and tried to run the ECU, but it 's an error saying : "The EISA config. utility is being executed in virtual 86 mode. Reboot and remove the installed memory manager before executing this utility again." What can i do ? Thank's, Matheus.
    Posted to SPM Digest (Forum) by Matheus on Wed, Dec 5 2012
  • Re: Nanoscope IIIa problem

    Hi, Case 1: The support of Bruker said this to me : We have a bad controller . DSP works as we have a line in scope window& as it doesn’t move A/D convertor boards have failed inside the controller. Unfortunately we don’t have parts to repair controllers (other than NS5 ). If the problem is in the circuit A/ D controller... Can I replace
    Posted to SPM Digest (Forum) by Matheus on Mon, Nov 26 2012
  • Nanoscope IIIa problem

    Hello guys, I'm having problems with a Multimode SPM Nanoscope IIIa and I need help to solve them. I have two computers to manage the system and both are having problems (different problems): Computer 1 : Communication error - Specs : Pentium 4 with an ISA DSP board and Windows XP/Nanoscope V5.31r3 - The problem : After a problem with the power
    Posted to SPM Digest (Forum) by Matheus on Tue, Nov 6 2012