Hi Darice, This post is 1 year old but my answer could help somebody else. I had the same problem while engaging with a very soft probe (DNP-D) in fluid in the PFQNM mode. the default engage step is 15 um, You will be able to engage with a lower engage step (microscope menu > engage settings), you could also use a lower gain. kind regards Celine
-
thanks for your answer Nic, I was wondering if you could help me with the choice of a standard sample. I found this one. Normal 0 false false false EN-AU X-NONE X-NONE MicrosoftInternetExplorer4 http://www.ntmdt-tips.com/products/view/tgx1 Celine
-
Hi All, According to the manual the MM8 is not designed to realise quantitative LFM. does anybody have an experience to share about manual calibration techniques? regards Celine
-
Hi Janne, Thank you for your useful advice. It worked well.
-
Hi, I am using the BioScope Catalyst. I can’t manage to engage in fluid with the DNP-10 probe (the “D” cantilever). The motor don’t move, I obtain a false engagement, even if I’m going deep in the liquid (to avoid any false engage due to the liquid surface). I have not this issue in air. I also tried to remove the “C”